Before the introduction of the CRO in its current form, the cathode ray tube had already been in use as a measuring device. The cathode ray tube is an evacuated glass envelope, similar to that in a black-and-white television set, with its flat face covered in a phosphorescent material (the phosphor). The screen is typically less than 20 cm in diameter, much smaller than the one in a television set.
In the neck of the tube is an electron gun, which is a heated metal plate with a wire mesh (the grid) in front of it. A small grid potential is used to block electrons from being accelerated when the electron beam needs to be turned off, as during sweep retrace or when no trigger events occur. A potential difference of at least several hundred volts is applied to make the heated plate (the cathode) negatively charged relative to the deflection plates. For higher bandwidth oscilloscopes where the trace may move more rapidly across the phosphor target, a positive post-deflection acceleration voltage of over 10,000 volts is often used, increasing the energy (speed) of the electrons that strike the phosphor. The kinetic energy of the electrons is converted by the phosphor into visible light at the point of impact. When switched on, a CRT normally displays a single bright dot in the center of the screen, but the dot can be moved about electrostatically or magnetically. The CRT in an oscilloscope uses electrostatic deflection.
Between the electron gun and the screen are two opposed pairs of metal plates called the deflection plates. The vertical amplifier generates a potential difference across one pair of plates, giving rise to a vertical electric field through which the electron beam passes. When the plate potentials are the same, the beam is not deflected.
When the top plate is positive with respect to the bottom plate, the beam is deflected upwards; when the field is reversed, the beam is deflected downwards. The horizontal amplifier does a similar job with the other pair of deflection plates, causing the beam to move left or right. This deflection system is called electrostatic deflection, and is different from the electromagnetic deflection system used in television tubes. In comparison to magnetic deflection, electrostatic deflection can more readily follow random changes in potential, but is limited to small deflection angles.
The timebase is an electronic circuit that generates a ramp voltage. This is a voltage that changes continuously and linearly with time. When it reaches a predefined value the ramp is reset, with the voltage reestablishing its initial value. When a trigger event is recognized the reset is released, allowing the ramp to increase again. The timebase voltage usually drives the horizontal amplifier. Its effect is to sweep the electron beam at constant speed from left to right across the screen, then quickly return the beam to the left in time to begin the next sweep. The timebase can be adjusted to match the sweep time to the period of the signal.
Meanwhile, the vertical amplifier is driven by an external voltage (the vertical input) that is taken from the circuit or experiment that is being measured. The amplifier has a very high input impedance, typically one megohm, so that it draws only a tiny current from the signal source. The amplifier drives the vertical deflection plates with a voltage that is proportional to the vertical input. Because the electrons have already been accelerated by hundreds of volt, this amplifier also has to deliver almost hundred volts, and this with a very high bandwidth. The gain of the vertical amplifier can be adjusted to suit the amplitude of the input voltage. A positive input voltage bends the electron beam upwards, and a negative voltage bends it downwards, so that the vertical deflection of the dot shows the value of the input. [3]
The response of this system is much faster than that of mechanical measuring devices such as the multimeter, where the inertia of the pointer slows down its response to the input.
When all these components work together, the result is a bright trace on the screen that represents a graph of voltage against time. Voltage is on the vertical axis, and time on the horizontal.
Observing high speed signals, especially non-repetitive signals, with a conventional CRO is difficult, due to non-stable or changing triggering threshold which makes it hard to "freeze" the waveform on the screen. This often requires the room to be darkened or a special viewing hood to be placed over the face of the display tube. To aid in viewing such signals, special oscilloscopes have borrowed from night vision technology, employing a microchannel plate in the tube face to amplify faint light signals.
Tektronix Model C-5A Oscilloscope Camera with Polaroid instant film pack back.
Although a CRO allows one to view a signal, in its basic form it has no means of recording that signal on paper for the purpose of documentation. Therefore, special oscilloscope cameras were developed to photograph the screen directly. Early cameras used roll or plate film, while in the 1970s Polaroid instant cameras became popular.
The vertical amplifier and timebase controls are calibrated to show the vertical distance on the screen that corresponds to a given voltage difference, and the horizontal distance that corresponds to a given time interval.
The power supply is an important component of the scope. It provides low voltages to power the cathode heater in the tube, and the vertical and horizontal amplifiers. High voltages are needed to drive the electrostatic deflection plates. These voltages must be very stable. Any variations will cause errors in the position and brightness of the trace.
Later analog oscilloscopes added digital processing to the standard design. The same basic architecture - cathode ray tube, vertical and horizontal amplifiers - was retained, but the electron beam was controlled by digital circuitry that could display graphics and text mixed with the analog waveforms. The extra features that this system provides include:
* on-screen display of amplifier and timebase settings;
* voltage cursors - adjustable horizontal lines with voltage display;
* time cursors - adjustable vertical lines with time display;
* on-screen menus for trigger settings and other functions.